ISO-STIM 01D
Features
- No batteries needed (external power supply)
- Isolated Output
- Three input modes: ANALOGUE, GATE TTL, DIRECT
- Stimulus AMPLITUDE set by a three digit control or by an external source
- Stimulus polarity set by switch (+, 0 ,-)
- Current (max. ±10 mA) or Voltage (max. ±100 V) source output
- „Overrange“ and „active“ LED indicators
Specifications
INPUT OPERATING MODES:
DIRECT, ANALOG TRIGGER, TTL TRIGGER, selectable with toggle switch
FUNCTION MODES:
voltage source, current source, selectable with toggle switch
SOURCE RANGES:
±1 mA [optional ±10 mA] (current source),
±100 V [optional ±45 V] (voltage source), selectable with toggle switch
TIME RANGE FACTORS:
×10 µs, ×100 µs, ×1ms, selectable with toggle switch, time range up to 99 ms
AMPLITUDE POTENTIOMETER:
3 digits, ±10 V, amplified internal with gain of 10 (voltage source), polarity selectable with toggle switch
TIME POTENTIOMETER:
2 digits, 0 - 99 as multiplication factor selectable
INPUT VOLTAGE RANGE:
±10 V, amplified internally with gain of 10 (voltage source) or gain of 1 (current source)
OUTPUT VOLTAGE:
±100 V max., isolated (optional ±45 V)
OUTPUT CURRENT:
±1 mA max., isolated (optional ±10 mA)