A-M Systems 2400

PrintShare
Amplifier for Voltage Clamping using Patch Electrodes on Single Channels or Whole Cells
AM Systems Model 2400 Patch Clamp Amplifer
Description

The Model 2400 is a low noise, full featured intracellular/extracellular amplifier designed for voltage clamping using patch electrodes on single channels or whole cells. Its unique design allows fast intracellular current clamp measurements with sharp electrodes. Advanced circuit design techniques using field programmable gate arrays eliminate noisy microprocessors.

Features

Specifications

Probe Gain and Bandwidth

Feedback Resistor

Probe Gain

Maximum Current

Minimum Bandwidth

10 MΩ

10 mV/nA

1000 nA

100 kHz

100 MΩ

100 mV/nA

100 nA

100 kHz

1 GΩ

1 mV/pA

10 nA

60 kHz

10 GΩ

10 mV/pA

1 nA

40 kHz

 

BNC input connector: 1 mm ground pin connector
The probe has three connections, the input BNC, a ground jack, and a guard jack.
The input BNC will hold any BNC style electrode holder, and is where any input signal should be connected. The center conductor is connected through a current limiting resistor to a field effect transistor amplifier. Other than the current limiting resistor there is no other protection to the probe so care must be taken to avoid electrostatic discharges. The outer conductor of the BNC is connected to a guard circuit. The guard is a buffered version of the input voltage, so care must be taken not to ground the outer conductor of the BNC.

Probe Case, Size: 54 mm long × 20 mm high × 36 mm wide

 

Maximum Instrument Noise

Voltage clamp modes measured with the internal 4 pole Bessel filter and minimum external noise

Bandwidth

Headstage Resistor

10 MΩ

50 MΩ

100 MΩ

500 MΩ

1 GΩ

5 GΩ

10 GΩ

50 GΩ

1 kHz Theoretical Limit

0.002 nA

0.6 pA

0.4 pA

0.2 pA

0.1 pA

0.06 pA

0.04 pA

0.02 pA

1 kHz

0.003 nA

0.7 pA

0.5 pA

0.5 pA

0.5 pA

0.5 pA

0.5 pA

0.5 pA

Headstage

The probe is both a current to voltage converter and a voltage follower. The current to voltage converter used in voltage clamping has one of two high value resistors used for amplifying membrane currents. Below is a table of the different feedback resistor configurations of A-M Systems probes you can purchase from us.

Note: For proper operation, the purchase of a Model 2400 Patch Clamp Amplifier headstage and an electrode holder is required (order separately).

Headstage Model 2400

 

 

Catalog Number

Resistor

Maximum Current

Probe High Gain
Resistor Value

Probe High Gain
Maximum Current

880210

10 MΩ

1000 nA

1 GΩ

10 nA

880218

10 MΩ

1000 nA

10 GΩ

1 nA

880222*

100 MΩ

100 nA

10 GΩ

1 nA

* Standard headstage delivered with Model 2400 Patch Clamp Amplifier unless one of the other catalog numbers is selected.