Microwire Arrays (MWA)

Implantierbare Mikroelektrodenträger, Platin/Iridium - MicroProbes
Beschreibung

 

Designed for chronic experiments, the MicroProbes Microwire Arrays (MWA) provide the optimum in flexibility and affordable alternative for the neuroscience community.

Features

  • Up to 32 penetrating wires per array for short- and long-term recordings.
  • Custom wire length (5-23 mm), including different length for each row.
  • Custom spacing between wires is in the range of 250 - 1000 µm.
  • Custom wire diameter options available (25/50 µm).
  • Custom impedance for individual wires between 200-800 kΩ.
  • Stainless steel and Platinum/Iridium fine wires with teflon or polyimide insulation material.
  • Custom reference and ground electrodes according to the researcher’s requirements.
  • Optionally integrated fiber optic and/or drug delivery cannula.

 

MWA Characteristics

Diameter / Configuration / Spacing
Microwire Array Characteristics

 

Impedance Values

Impedance values per Metal type and Wire size

Metal type Wire thickness Impedance
Pt/Ir 25 µm 0.4 to 0.8 MΩ
Pt/Ir 50 µm 0.2 to 0.4 MΩ
Stainless Steel 25 µm 0.6 to 0.9 MΩ
Stainless Steel 50 µm 0.4 to 0.6 MΩ



 

Publications:

Axial levodopa-induced dyskinesias and neuronal activity in the dorsal striatum
Levodopa-induced dyskinesias are abnormal involuntary movements that limit the effectiveness of...

Corticostriatal Field Potentials Are Modulated at Delta and Theta Frequencies during Interval-Timing Task in Rodents
Organizing movements in time is a critical and highly conserved feature of mammalian behavior...

D1-dependent 4 Hz oscillations and ramping activity in rodent medial frontal cortex during interval timing
Organizing behavior in time is afundamental process that is highly conserved across species.Here...

Electrophysiological evidence of alterations to the nucleus accumbens and dorsolateral striatum during chronic cocaine self-administration
As drug use becomes chronic, aberrant striatal processing contributes to the development of...

Infusion of D1 Dopamine Receptor Agonist into Medial Frontal Cortex Disrupts Neural Correlates of Interval Timing
Medial frontal cortical (MFC) dopamine is essential for the organization of behavior in time. Our...

 

 

How to order:

MicroProbes
Please contact the manufacturer, MicroProbes directly for further specifications and ordering.  We would appreciate if you mention "Science Products" at any order or inquiry.

 

Online design form

MWA - Microwire Arrays

 

 

 

 

Related Products:

MicroProbes FMA

Floating Microelectrode Arrays

(MicroProbes)
 

MicroProbes LMA

Linear Microelectrode Arrays

(MicroProbes)

MicroProbes MBA

Microwire Brush Arrays

(MicroProbes)

MicroProbes MEA

Microelectrode Arrays

(MicroProbes)

MicroProbes MWA

Microwire Arrays

Pt/Ir (MicroProbes)

MicroProbes Opto-MEA

Optogenetic Arrays / MEA

Pt/Ir (MicroProbes)

MicroProbes Opto-MWA

Optogenetic
Mircowire Arrays / MWA

Pt/Ir (MicroProbes)

 

MicroProbes Electrode Arrays

Array Comparison Chart

(PDF)